Model
|
COD-9100
|
|
Measuring Mode
|
COD / ABS / Temperature
|
|
COD Range
|
0~200 mg/L, 5~1000 mg/L, 50~2500mg/L, 100~5000 mg/L, ( 500~10,000 mg/L optional)
Different ranges can be measured by changing the ratio of reagent to the water sample
|
|
Resolution
|
0.1 mg/L
|
|
Accuracy
|
<100 mg/L: <±10 mg/L
>100 mg/L: <±10 % of measuring values
Measuring accuracy can be increased by raising the water sample ratio
|
|
Repeatability
|
<±5% of measuring values
|
|
Principle
|
Closed Rapid Digestion Spectrophotometry---Potassium Dichromate method
|
|
Ambient Conditions
|
Temperature: 5~50°C; Humidity: <90% (without condensation)
|
|
Display
|
10 inch of HMI with color touch screen display
Transmitter with large LCM display
|
|
Languages
|
English/ Traditional Chinese/ Simplified Chinese
|
|
Digestion Time
|
3 ~ 120 minutes (Freely selectable)
|
|
Min. Measuring Time
|
Approx. 30 minutes (as digestion time is set as 10 minutes)
|
|
Measuring Cycle
|
Fixed duration of 25~9999 minutes, or self-defined on the hour
|
|
Calibration
|
Manual or automatic calibration, automatic calibration duration: 1~99 days
|
|
Contact Output
|
Relay 1 & 2
|
RELAY ON/OFF contact, 240VAC 0.5A Max. (recommended)
|
|
Two individual programmable contacts for Hi point, Lo point, measurement, calibration, wash and error shutdown
|
|
Relay 3
|
RELAY ON/OFF contact, 250VAC 10A Max. (recommended)
|
|
One control point for sampling pump
|
|
Analog Output 1
|
Isolated DC 0/4~20mA corresponding to COD measurement, max load 500Ω
|
|
Analog Output 2
|
Isolated DC 0/4~20mA corresponding to COD measurement or temperature, max load 500Ω
|
|
Digital Interface
|
HMI
|
Isolated RS-485 interface, with Modbus RTU protocol, offers readouts, calibration data, setup information & error message, and available for remote setting of control parameters
|
|
Transmitter
|
Isolated RS-485 interface, timing interval outputs COD measurement, temperature, status & time
|
|
HMI
|
- Remote control via Modbus 2. Detachable and controllable via RS-485
|
|
Transmitter
|
- Running programs of transmitter are regulated by HMI
- While not connected to HMI, works independently based on set programs
- Directly trigger measurement or emergent shutdown via transmitter under auto or
measurement status
|
|
Reagent
Depletion
|
One sets of reagents( 1 liter oxidant, 1 liter masking agent, 1 liter catalyst) for proceeding about 500 measurements (Changes depend on measuring cycles and amounts of reagent)
|
|
Piping/solenoid valve
|
Acid-resistant PFA material applied on all the path reagent goes through
|
|
Data Logger
|
HMI records measurement data, calibration data and warning data up to 10,000 data sets
Download data via USB-A interface
|
|
Power Supply
|
Specify single-phase power voltage and frequency when ordering
|
|
01: 220VAC(-15~+10%)/60Hz
|
02: AC 220VAC(-15~+10%)/50Hz
|
|
Power
|
275W Max
|
|
Dimensions
|
163cm × 53cm × 46cm (H×W×D)
|
|
Weight
|
Approx. 66 Kg (without reagents)
|
|
Protection
|
Human Machine Interface: IP65 (NEMA 4X)
|