4-Electrode Design
The 8-24X series conductivity sensors apply the state of art 4-Electrode measurement principle to accurately measure medium to high conductivity solutions. Its modular cable and sensor design provides the most convenience of sensor re-installation. The 8-241-01 sensor is a HF resistant one, while the 8-242 sensor can stand higher temperature up to 120˚C.
|
Order number: |
8-241 |
8-241-01 |
8-242 |
|
Type |
4-electrode sensor |
||
|
Range: |
20.00μS~500.0mS |
||
|
Cell constant |
0.5 cm-1 |
||
|
Working temp. |
0~70°C |
0~70°C |
0~120°C |
|
Pressure |
0~7 bar at 25°C(77°F) |
||
|
HF Resistant |
- |
Yes |
- |
|
Electrode Material |
Titanium, CPVC |
Monel, CPVC |
Titanium, PVDF |
|
Body Material |
CPVC |
CPVC |
PVDF |
|
Temp. sensor |
NTC 30K |
||
|
Insertion length |
40mm |
||
|
Max. length |
120mm |
||
|
Process connection |
3/4”NPT |
||
|
Compatible cable length: 5m (8-105), 10m (8-110), 20m(8-120), 30m(8-130) |
|||
|
Compatible bypass chamber: 8-TF02, Compatible T-Connector: 8-TF01 |
|||